(a) How does the amperometric glucose monitor in Figure 7-12 work? b) Why is a mediator advantageous in the glucose monitor? c) How does the coulometric glucose monitor in Figure 17-14 work? (d) Why does the signal in the amperometric measurement depend on the temperature of the blood sample, whereas the signal in coulometry is independent of temperature? Do you expect the signal to increase or decrease with increasing temperature in amperometry?

(e) Glucose(C6H12O6$,FM180.16)is present in normal human blood at a concentration near 1g/L How many microcoulombs are required for complete oxidation of glucose in 0.300μL of blood in a home glucose monitor if the concentration is 1.00g/L ?

Short Answer

Expert verified

(a)When a drop of blood is placed on the strip, the glucose in the blood is transformed to gluconolactone, lowering the mediator. When a voltage of is applied to the carbon electrode, re-oxidation in the decreased mediator happens. The cureent affects the rate of oxidation of the mediator between the two electrodes. The rate at which mediators are oxidised is proportional to the amount of glucose and other metabolites in the bloodstream.

(b) Because the mediator's concentration is constant, fluctuations in electrode current due to changes in glucose concentration are unaffected. The interference of other species in blond is reduced by lowering the electrode voltage of the oxidation mediator.

(c)(c) A coulometric sensor can calculate the number of electrons required to oxidise glucose in a blood sample.

(d)(d) The blood temperature rises as the current increases. The number of electrons emitted during oxidation is measured using coulometry. Independent of temperature, glucose emits two electrons per molecule. Furthermore, the coulometric signal is unaffected by temperature.

(e) (e) The amount of current required to maintain constant glucose levels is

Q=nF=left3.33times10-9right96485C/mol=321muC

Step by step solution

01

Concept used

It is necessary to discuss the method of using an amperometric glucose monitor to monitor glucose levels.

02

Step 2:

(a)

The procedure of using an amperometric glucose monitor to measure glucose levels.

A test strip with two carbon electrodes and a reference electrode is included in an amperometric glucose monitor (silver-silver chloride). The carbon electrode is covered with a mediator and glucose oxidase. When a drop of blood is placed on the strip, the glucose in the blood is oxidised to gluconolactone, and the mediator is decreased. When a potential of is applied to the carbon electrode, re-oxidation in decreased mediator occurs. Between both electrodes, the rate of oxidation of the mediator is precisely proportional to the cureent. The rate of oxidation of mediator is proportional to the strength of glucose and other species in blood. The second indicator electrode contains simply mediator and no glucose oxidase. Current measurement between two points.

03

Step 3:

(b)

The advantages of using a glucose monitor with a mediator.

The oxidation rate of glucose, which is dependent on the quantity of oxygen in the blood, is affected by the absence of mediator. Because of the low current, the monitor displays a low glucose concentration when the oxygen concentration drops. A mediator, such as dimethylferrocene, can replace oxygen in the oxidation of glucose, resulting in a reduction at the indicator electrode. Because the concentration of the mediator is constant, changes in electrode current owing to changes in glucose concentration are unaffected. The interference of other species in blond is reduced by lowering the electrode voltage of the mediator for oxidation.

04

Step 4:

(c)

The procedure for using a coulnmetric glucose monitor to measure glucose levels.

Glucose oxidase, which replaces glucose dehydrogenase, does not need oxygen as a reactant. It oxidises glucose by reducing the PQQ cofactor to PQQH2.

By oxidising the PQQH2to PQQ the polymer chain creates PQQ .Electrons are exchanged fromOs2+ to Os3+It can reach the carbon electrode through electron migration. The coulometric sensor can calculate the amount of electrons necessary to oxidise glucose in a blood sample.

05

Step 5:

(d)

The signal's amperometric measurement dependent.

Amperometry is used to quantify the current generated when glucose is oxidised by an enzyme. The rate of oxidation is proportional to the amount of current flowing. The rate of a chemical reaction is proportional to its temperature in general. As a result, current can raise the blood temperature as it rises. The number of electrons released during oxidation is determined via coulometry. Glucose, regardless of temperature, expels two electrons per molecule. Furthermore, the coulometric signal is unaffected by temperature.

06

Step 6:

(e)

A test strip with two carbon electrodes and a reference electrode is included in an amperometric glucose monitor (silver-silver chloride). The carbon electrode is covered with a mediator and glucose oxidase. When a drop of blood is placed on the strip, the glucose in the blood is oxidised to gluconolactone, and the mediator is decreased. When a potential of 0.2V is applied to the carbon electrode, re-oxidation in decreased mediator occurs. Between both electrodes, the rate of oxidation of the mediator is precisely proportional to the cureent. The rate of oxidation of mediator is proportional to the strength of glucose and other species in blood. The second indicator electrode contains simply mediator and no glucose oxidase. The strength of the species other than the reference electrode and the indication electrode two is used to measure the current between them.

The current required to monitor glucose levels at a fixed concentration is321μC

To calculate: The amount of current needed to keep track of glucose levels at a certain concentration.

1.00g of glucose = 5.55mMGlicose

0.300×10-6Lhas 1.665nmol glucose

During oxidation, glucose releases two electrons per molecule, regardless of temperature.

2×1.665nmol=3.3nmolelectrons are liberated.

Q=nF=left3.33times10-9right96458C/mol=321muC

Unlock Step-by-Step Solutions & Ace Your Exams!

  • Full Textbook Solutions

    Get detailed explanations and key concepts

  • Unlimited Al creation

    Al flashcards, explanations, exams and more...

  • Ads-free access

    To over 500 millions flashcards

  • Money-back guarantee

    We refund you if you fail your exam.

Over 30 million students worldwide already upgrade their learning with Vaia!

One App. One Place for Learning.

All the tools & learning materials you need for study success - in one app.

Get started for free

Most popular questions from this chapter

For a rotating disk electrode operating at sufficiently great potential, the redox reaction rate is governed by the rate at which analyte diffuses through the diffusion layer to the electrode (Figure 17-15b). The thickness of the diffusion layer is

δ=1.61D1/3V1/6ω-1/2

whereis the diffusion coefficient of reactant (m2/s),vis the kinematic viscosity of the liquid

(=viscosity/debsity=(m2/s),v and ωis the rotation rate (radians/s) of the electrode. There are 2π radians in a circle. The current densityis localid="1655441451764" (A/m2)is

localid="1655441445229" Currentdebsity=0.62nFD2/3v-1/6ω1/2C0

where nis the number of electrons in the half-reaction, Fis the Faraday constant, and localid="1655441459070" C0is the concentration of the electroactive species in bulk solution localid="1655441466748" (mol/m3,notmol/L) Consider the oxidation oflocalid="1655441474339" Fe(CN)64-in a solution of localid="1655441479067" 10.0mMK3Fe(CN)6+50.0mMK4Fe(CN)6at +0.90V(versus S.C.E.) at a rotation speed oflocalid="1655441490849" role="math" 2.00×103revolutions per minute. 27The diffusion coefficient oflocalid="1655441497131" Fe(CN)64-is2.5×10-9m2/sand the kinematic viscosity islocalid="1655441503345" 1.1×10-5m2/sCalculate the thickness of the diffusion layer and the current density. If you are careful, the current density should look like the value in Figure 17-16b.

Propagation of uncertainty. In an extremely accurate measurement of the Faraday constant, a pure silver anode was oxidized to Ag1 with a constant current of 0.2036390(60.0000004)A for18000.075(60.010s to give a mass loss of4.097900(60.0000003) g from the anode. Given that the atomic mass of Ag is 107.8682 (60.0002), find the value of the Faraday constant and its uncertainty.

Explain how the endpoint is detected in a Karl Fischer titration in Figure 17-35.

Chlorine has been used for decades to disinfect drinking water. An undesirable side effect of this treatment is reaction with organic impurities to create organochlorine compounds, some of which could be toxic. Monitoring total organic halide (designated TOX) is required for many water providers. A standard procedure for TOX is to pass water through activated charcoal, which adsorbs organic compounds. Then the charcoal is combusted to liberate hydrogen halides:

Organichalide(RX)o2/800°CCO2+H2O+HX

HX is absorbed into aqueous solution and measured by coulometric titration with a silver anode:

X-(aq)+Ag(s)AgX(s)+e-

When 1.00 L of drinking water was analyzed, a current of 4.23 mA was required for 387 s. A blank prepared by oxidizing charcoal required 6 s at 4.23 mA. Express the TOX of the drinking water as μmol halogen/L. If all halogen is chlorine, express the TOX as μgCI/L.

17-16. The sensitivity of a coulometer is governed by the delivery of its minimum current for its minimum time. Suppose thatcan be delivered for 0. 1s.

(a) How many moles of electrons are delivered by 5mAfor 0. 1s ?

(b) How many millilitres of a 0.01 Msolution of a two-electron reducing agent is required to deliver the same number of electrons?

See all solutions

Recommended explanations on Chemistry Textbooks

View all explanations

What do you think about this solution?

We value your feedback to improve our textbook solutions.

Study anywhere. Anytime. Across all devices.

Sign-up for free