A reversed-phase separation of a reaction mixture calls for isocratic elution with 48%methanol 52%water. If you want to change the procedure to use acetonitrile/water, what is a good starting percentage of acetonitrile to try?

Short Answer

Expert verified

Considering a vertical line (intersects each solvent line at the same elegant strength) at 48% methanol intersects the acetonitrile at 38%. Use acetonitrile/water instead of 48% methanol water 52%, the good starting percentage of acetonitrile must be 38%.

Step by step solution

01

Definition of isocratic elution

When the mobile phase has a constant concentration, isocratic elution is a term used in chromatography where the concentration of the mobile phase is constant throughout the chromatographic process.

02

Percentage of acetonitrile

Considering a vertical line (intersects each solvent line at the same elegant strength) at 48% methanol intersects the acetonitrile at 38%. Use acetonitrile/water instead of 48% methanol water 52%, the good starting percentage of acetonitrile must be 38%.

Unlock Step-by-Step Solutions & Ace Your Exams!

  • Full Textbook Solutions

    Get detailed explanations and key concepts

  • Unlimited Al creation

    Al flashcards, explanations, exams and more...

  • Ads-free access

    To over 500 millions flashcards

  • Money-back guarantee

    We refund you if you fail your exam.

Over 30 million students worldwide already upgrade their learning with Vaia!

One App. One Place for Learning.

All the tools & learning materials you need for study success - in one app.

Get started for free

Most popular questions from this chapter

The figure shows the separation of two enantiomers on a chiral stationary phase.


Sepation of enantiomers of Ritalin by HPLC with a chiral stationary phase.[Data from R.Bakhitar,L.Ramos,and F.L.S.Tse, “Quantification of methlylphenidate in plasma using chiral Liquid-chromatography/Tandem mass spectrometry: Application to Toxicokinetric studies,”Anal Chim Acta 2002,469,261.]

(a)From trandw1/2 find N for each peak.

(b) Fromtrandw1/2find the resolution.

(c)Giventm=1.62min, use Equation23-23with the average N to predict the resolution.

A mixture of 14compounds was subjected to a reversed-phase gradient separation going from 5%to 100%acetonitrile with

a gradient time of 60min. The sample was injected at t =time. All peaks were eluted between 22and 50min.

(a) Is the mixture more suitable for isocratic or gradient elution?

(b) If the next run is a gradient, select the starting and ending %acetonitrile

and the gradient time.

Question:The graph shows retention data from aC8silica column with an acetonitrile/water mobile phase.


(a) What mobile phase composition provides greatest retention(k)for the components? Least retention? Coelution (equalk)of two components?

(b) Predict the retention time of each peak at 40%and60%acetonitrite. Draw a chromatogram (a "stick diagram" representing each peak as a vertical line) of the separation at each mobile phase composition.

(c) Would60%acetonitrile yield adequate resolution?

(d) Assuming Gaussian peaks, does the separation at60%acetonitrile have the attributes of a good separation?

a. Why are HPLC particles porous?

b. Why are the particles with60-120A°pores used for small molecules but wide pore300-A°stationary phases used to separate polypeptides and proteins?s

In hydrophilic interaction chromatography (HILIC), why is eluent strength increased by increasing the fraction of water in the mobile phase?

See all solutions

Recommended explanations on Chemistry Textbooks

View all explanations

What do you think about this solution?

We value your feedback to improve our textbook solutions.

Study anywhere. Anytime. Across all devices.

Sign-up for free