Two peaks emerge from a reversed-phase chromatography column as sketched in the illustration.

According to Equation 23-33, resolution is given by

Resolution=N4(α-1)α(k21+k2)

where Nis plate number, αis relative retention (Equation23-20), and k2 is the retention factor for the more retained component (Equation 23-16).

(a) If you decrease the amount of organic solvent in the mobile phase, you will increase retention. Sketch the chromatogram if retention factors increase but Nand αare constant.

(b) If you change the solvent type or the stationary phase, you will change the relative retention. Sketch the chromatogram ifαincreases but Nandk1are constant.

(c) If you decrease particle size or increase column length, you can increase the plate number. Sketch the chromatogram if Nincreases by (i) decreasing particle size and (ii) increasing column length. Assume αand k2are constant.

Short Answer

Expert verified

a)The sketch of chromatogram if retention factors increase but Nand αare constant:

b) The sketch of chromatogram if αincreases but Nand k1are constant:

c) The sketch of chromatogram if Nincreases by (i) decreasing particle size and (ii) increasing column length.

Step by step solution

01

Definition of retention factor.

The retention factor (Rf) is used in thin-layer chromatography to compare and identify substances. The distance travelled by the compound divided by the distance travelled by the solvent front gives the Rf value of a compound.

02

Step 2: The sketch of chromatogram if retention factors increase but   and  are constant.

a)

When the quantity of organic solvent in the mobile phase is reduced, the retention is automatically enhanced. As a result, ifNandαare held constant, the chromatogram will look like this

03

Step 3: The sketch of chromatogram if αincreases butNandk1are constant.

b)

If the solvent or stationery phase is changed, the relative retention will vary. As a result, the chromatogram will look like this:

04

Step 4: The sketch of chromatogram if N increases by (i) decreasing particle size and (ii) increasing column length.

c)

The chromatogram can be produced as shown below if the particle size is reduced and the plate number is increased.

The chromatogram can be obtained as shown below if the column length is raised and the plate number is increased.

Unlock Step-by-Step Solutions & Ace Your Exams!

  • Full Textbook Solutions

    Get detailed explanations and key concepts

  • Unlimited Al creation

    Al flashcards, explanations, exams and more...

  • Ads-free access

    To over 500 millions flashcards

  • Money-back guarantee

    We refund you if you fail your exam.

Over 30 million students worldwide already upgrade their learning with Vaia!

One App. One Place for Learning.

All the tools & learning materials you need for study success - in one app.

Get started for free

Most popular questions from this chapter

What are the general steps in developing as isocratic separation for reversed-phase chromatography with one organic solvent and temperature as variable?

Why are the relative eluent strengths of solvents in adsorption chromatography fairly independent of solute?

In monolithic columns60 the stationary phase is a single porous piece of silica or polymer filling the entire column and synthesized within the column from liquid precursors. Monolithic columns offer similar plate height to HPLC particles, but with less resistance to flow. Therefore, faster flow or longer columns can be used. The figure shows separation of isotopic molecules on a long monolithic column. Packed columns have too much resistance to flow to be made so long.

Separation of isotopic molecules on a 440-cm-long monolithic C18-silica column eluted withCH3CN/H2O(30: 70 vol/vol) at 308C. [Data from K. Miyamoto, T. Hara, H. Kobayashi, H. Morisaka, D. Tokuda, K. Horie, K. Koduki, S. Makino, O. Nuñez, C. Yang, T. Kawabe, T. Ikegami, H. Takubo, Y. Ishihama, and N. Tanaka, “High-Efficiency Liquid Chromatographic Separation Utilizing Long Monolithic Silica Capillary Columns,” Anal. Chem. 2008, 80, 8741.]

(a) Unretained thiourea is eluted in 41.7 min. Find the linear velocity ux (mm/s).

(b) Find the retention factor k forC6D6

(c) Find the plate number N and plate height forC6D6

(d) Assuming that the peak widths forC6H5Dand C6D6are the same as that of C6D6, find the resolution of C6H5Dand C6D6.

(f) If we just increased the column length to increase N, what value of N and what column length would be required for a resolution of 1.000?

(g) Without increasing the length of the column, and without changing the stationary phase, how might you improve the resolution?

(h) When the solvent was changed fromCH3CN/H2O(30:70 vol/vol) toCH3CN/CH3OH/H2O(10:5:85), the relative retention for C6H5D andC6D6increased to 1.0088 and the retention factor for C6H6 changed to 17.0. If the plate number were unchanged, what would be the resolution?

Chromatography–mass spectrometry. HPLC separation of

enantiomers of the drug Ritalin on a chiral stationary phase was

shown in Problem 25-13.

(a) Detection is by atmospheric pressure chemical ionization with

selected reaction monitoring of the m/z 23484 transitions. Explain

how this detection works and propose structures for m/z 234 and m/z 84.

(b) For quantitative analysis, the internal standardH32-Ritalin with

a deuterated methyl group was added. Deuterated enantiomers have

the same retention times as unlabelled enantiomers. Which selected

reaction monitoring transition should be monitored to produce a

chromatogram of the internal standard in which unlabelled Ritalin

will be invisible?

Literature search problem: The purity of cocaine bought on the street varies dramatically. Cutting agents include levamisole, a compound normally used to kill parasites. Search the literature for a reversed-phase liquid chromatography method with diode array detection for the determination of the purity of street cocaine.

(a) Give the citation (authors, title, journal name, year, volume, pages) for the research paper that fits the criteria of this analysis.

(b) What alternative methods could be used for analysis of adulterated street cocaine?

(c) What is the range of purity of street cocaine?

(d) What type of analytical column was used for the analysis?

(e) How was the method validated?

(f) What is “robustness of an analytical method?”